Meter repairs
January 2024
7. Insert the transducer capsule assembly into the transducer stalk while aligning the
8. Ensure that the indicator ring on the transducer capsule assembly is flush with the
9. Screw the transducer retainer on the stalk and use a 1 ⅛-in. wrench to bottom out.
10. Connect the transducer chordset to the transducer retainer. The internal connector
11. Secure the transducer cabling nut by turning clock wise. Ensure the cable nut
12. Repeat
13. Continue with
60
anti-rotation pin on the capsule with the slot on the inner surface of the stalk.
Figure 3-15: Length verification of T-200 transducer capsule assembly
A. Transducer stalk
B. Indicator ring
C. Capsule assembly
end of the transducer stalk (see
the correct length.
Note
To avoid damages to the anti-rotation pins, do not torque down the transducer
retainer.
of the transducer chordset is keyed and will only go on one way.
threads are correctly aligned.
Step 1
through
Step 11
capsule assemblies were replaced.
Modifying the calibration parameters for T-Slot transducers
MeterLink Transducer Swap-out Wizard to update delta delay time.
Maintenance and Troubleshooting Manual
Figure
3-15). This will ensure the capsule is set to
for other transducer assemblies if their transducer
00809-0600-3104
to use the
Emerson.com/Rosemount