Diagnostic functionality

Heartbeat Technology

Petroleum

Petroleum & Product
identification
82

Application packages

Many different application packages are available to enhance the functionality of the device. Such
packages might be needed to address safety aspects or specific application requirements.
The application packages can be ordered with the device or subsequently from Endress+Hauser.
Detailed information on the order code in question is available from your local Endress+Hauser sales
center or on the product page of the Endress+Hauser website: www.endress.com.
Detailed information on the application packages:
Special Documentation →  87
Order code for "Application package", option EA "Extended HistoROM"
Comprises extended functions concerning the event log and the activation of the measured value
memory.
Event log:
Memory volume is extended from 20 message entries (standard version) to up to 100 entries.
Data logging (line recorder):
• Memory capacity for up to 1000 measured values is activated.
• 250 measured values can be output via each of the 4 memory channels. The recording interval can
be defined and configured by the user.
• Measured value logs can be accessed via the local display or operating tool e.g. FieldCare,
DeviceCare or Web server.
For detailed information, see the Operating Instructions for the device.
Order code for "Application package", option EB "Heartbeat Verification + Monitoring"
Heartbeat Verification
Meets the requirement for traceable verification to DIN ISO 9001:2008 Chapter 7.6 a) "Control of
monitoring and measuring equipment".
• Functional testing in the installed state without interrupting the process.
• Traceable verification results on request, including a report.
• Simple testing process via local operation or other operating interfaces.
• Clear measuring point assessment (pass/fail) with high total test coverage within the framework
of manufacturer specifications.
• Extension of calibration intervals according to operator' s risk evaluation.
Heartbeat Monitoring
Continuously supplies data, which are characteristic of the measuring principle, to an external
condition monitoring system for the purpose of preventive maintenance or process analysis. These
data enable the operator to:
• Draw conclusions - using these data and other information - about the impact the measuring
application has on the measuring performance over time.
• Schedule servicing in time.
• Monitor the process or product quality, e.g. gas pockets.
For detailed information, see the Special Documentation for the device.
Order code for "Application package", option EJ "Petroleum"
The most important parameters for the Oil & Gas Industry can be calculated and displayed with this
application package.
• Corrected volume flow and calculated reference density in accordance with the "API Manual of
Petroleum Measurement Standards, Chapter 11.1"
• A temperature measurement is required to calculate the corrected volume. The measured values
can be read in via the 4-20 mA input on the device, for example.
As a temperature measuring device, the resistance thermometer TST602 is recommended. TMT82
is recommended for use in hazardous areas.
For detailed information, see the Special Documentation for the device. →  87.
Order code for "Application package", option EQ "Petroleum& Product identification"