  • ENGLISH, page 1
 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
 This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section.
 The cuffs are sensitive and must be handled with care.
 Only pump up the cuff once fitted.
 Never inflate beyond 300 mmHg.
 Always deflate the cuff completely before storage.
 Do not use this device if you think it is damaged or notice
anything unusual.
 Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
 The measurement results given by this device is not a diag-
nosis. It is not replacing the need for the consultation of a physi-
cian, especially if not matching the patient's symptoms. Do not
rely on the measurement result only, always consider other
potentially occurring symptoms and the patient's feedback.
Calling a doctor or an ambulance is advised if needed.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.
Contra-indications
Do not use this device if the patient's condition meets the following
contra-indications, to avoid inaccurate measurements or injuries.
 Do not use this device in a moving vehicle (for example in a car
or on an aircraft).
 Contraindicated for neonate use. Do not use with neonatal cuffs
or neonate patients.
 The aneroid sphygmomanometer requires frequent recalibra-
tion (at least every 6 months).
 Expertise and retraining required to avoid observer error.
 It helps the user detect the Korotkoff sounds through a stetho-
scope for auscultation. The measurements and readings might
be interfered by the noise around.
 Requires manual dexterity to ensure proper cuff deflation rate.
 Excellent hearing and vision are required for accurate measu-
ments.
4

Side effects

Use of the device may be accompanied by minor side effects.
 A device can lose calibration (become inaccurate) when it is
jostled or bumped, leading to false readings.
 Observers might have bias and terminal digit preference which
lead to inaccurate readings.

Device care

Clean the device only with a soft, dry cloth.

Cleaning the cuff

Carefully remove spots on the cuff with a damp cloth and soap-
suds.
WARNING: Do not wash the cuff in a washing machine or
dishwasher!

Accuracy test

We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test (see foreword).

6. Guarantee

This device is covered by a 2 year guarantee from the date of
purchase. During this guarantee period, at our discretion, Microlife
will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
 Transport costs and risks of transport.
 Damage caused by incorrect application or non-compliance
with the instructions for use.
 Damage caused by accident or misuse.
 Packaging/storage material and instructions for use.
 Regular checks and maintenance (calibration).
 Accessories and wearing parts: Tubes, pumping ball, cuff.
Should guarantee service be required, please contact the dealer
from where the product was purchased, or your local Microlife
service. You may contact your local Microlife service through our
website: www.microlife.com/support
Compensation is limited to the value of the product. The guarantee
will be granted if the complete product is returned with the original
invoice. Repair or replacement within guarantee does not prolong
or renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.