Status - description

The status of the unit is indicated by description.
Figure 6 – Status - description
Program name
Program user
Priority
Current segment
Current loop
Status
Time set
Time elapsed
Time remaining
Status – protection and alarms
In the left part there is information about the alarm class and the upper set value. It is associated with a running or completed
program. The protection parameters can be set in the program parameters
protection classes
(12. Temperature
In the right part there is information about the set alarms (more information on alarms set, see
for overrunning up and down. The value 0 means the alarm is turned off.
Figure 7 - Status – protection and alarms
21
vwr.com I Instruction manual VWR
The name of program
The name of user to whom the program belongs
In regard of time or parameters, more information see
Currently performing segment / total number of segments in the program
Currently performing cycle/ total number of cycles to perform
Status of the program: ramp, set temperature, program off
Set time of performing segment
Elapsed time after reaching the segment
Remaining time until the end of the segment
protection).
®
CO
incubator
2
10.1.5. Priority
(10.1.1. Program creating /
(10.11.
10.1.4. Loop
edition). Information about
Alarms). Separately