VWR ILCO 180 Premium Instruction Manual page 16

Incu-line premium co2 incubator
Table of Contents

Advertisement

The gas connections must be installed by qualified personel who are trained in handling
the respective gases and familiar with the required safety measures.
Opening the gas cylinder valves is only permitted when the cylinder has been connected to the chamber.
Pressure level is checked by the controller regularly. When the CO
("CO2 Pressure low Start").
Open the gas cylinder valve slowly to avoid pressure surges.
Always close the valve even with apparently empty cylinders; screw on the cap when not in use.
Return gas cylinders with the valve closed.
Intended use
INCU-Line Premium CO
o
37
C. The chamber permits setting an exact CO
to avoid osmolarity increasing.
The appliance may only be loaded with materials and substances which cannot explode, burst or ignite (do not form any toxic
or explosive vapours at the set temperature). Potentially explosive gas-air mixtures must not form, neither in the working chamber
nor in the direct vicinity of the appliance. The incubator should not be used to sterilize samples. The sterilisation programme in
the menu will only sterilize the inner chamber.
The chamber is not classified as a medical device according to Medical Device Directive 93/42/EEC. Do not use it to sterilise
medical devices.
Symbols
The following chart is an illustrated glossary of the symbols that are used in this manual.
CAUTION This symbol indicates a potential risk and alerts you to proceed with caution.
CAUTION This symbol indicates important information and the recommendations to follow.
CAUTION This symbol indicates risks associated with hot surfaces.
This symbol means that a given field / window can be moved in the direction shown
in the picture.
16
vwr.com I Instruction manual VWR
incubators are intended for incubation of cell cultures and similar under typical conditions of approx.
2
®
CO
incubator
2
atmosphere inside and provides high humidity inside the chamber
2
pressure is low, an alarm occurs on the screen
2

Advertisement