Rs-485 Modbus™ Protocol; Table 3: Modbus™ Registers - Teledyne 700 IR Series Instruction Manual

Infrared gas detector for combustible and co2 gases
Table of Contents

Advertisement

4
RS-485 Modbus™ Protocol
OLCT 700/710 detectors feature Modbus™ compatible communications protocol and are addressable
via the program mode. Communication is two wire, half duplex 485, 9600 baud, 8 data bits, 1 stop bit,
no parity, with the detector set up as a slave device. A master controller up to 1200 meters (4000 feet)
away can theoretically poll up to 256 different detectors. This number may not be realistic in harsh
environments where noise and/or wiring conditions would make it impractical to place so many devices
on the same pair of wires. If a multi-point system is being utilized, each detector should be set for a different
address. Typical address settings are: 01, 02, 03, 04, 05, 06, 07, 08, 09, 0A, 0B, 0C, 0D, 0E, 0F, 10,
11, etc.
Detector RS-485 ID numbers are factory default to 01. These can be changed in the field via the Operator
Interface described in Section 3.5.5 Set Serial ID.
The following section explains the details of the Modbus™ protocol that the OLCT 700/710 detector
supports.
Code 03 - Read Holding Registers is the only code supported by the transmitter. Each transmitter contains
6 holding registers which reflect its current status.
FC
REG
Content Description
03
40000
Device Type
03
40001
Read Detectable Range
06
40001
Write Detectable Range
03
40002
Read Concentration
03 06
40003
Read AutoSpan Level
40003
Write AutoSpan Level
03
40004
Read Sensor Life
03
40005
Read Fault Status Bits
03
40006
Read Model #
03
40007
Read Days Since Cal
03
40008
4-20 Current Output
mA x100
03
40009
Read Input Voltage V x100
NP700IRIMEN
Revision E.0
INFRARED GAS DETECTOR FOR COMBUSTIBLE AND CO
Table 3: Modbus™ Registers
R/W
Value
R
8
OLCT 700/710 Detector
1,2
R/W
100
For 0-100
10000
For 0-10000
3,2
R
1000
Bound by range. If > range, this value is in
fault.
4,2
R/W
50
Span gas at 50
R
85
For 85% sensor life
5
R
0x0001
Global Fault
0x0002
Auto Span Fault
0x0004
Temperature Fault
0x0008
4-20mA Fault
0x0010
Input Voltage Fault
0x0020
Memory Fault
0x0040
Processor Fault
0x0080
Clearing Fault
0x0100
Stability Fault
0x0200
Range Fault
0x0400
Sensor Fault
0x0800
Zero Fault
0x1000
Sensor Fault 2
0x2000
<reserved>
0x4000
In Calibration
0x8000
Communication Error
R
1, 2, 3,
DM, FP, IR, TP, PID respectively
4, 5
R
29
29days
R
400
4.00mA
R
2400
24.00V
700 IR Series
Content Definition
Meaning
FP – Read only
2
TP – 20, 50, 100, 200
IR – 0 to 10000
FP – 5% to 95% of Range
(40001)
TP – 2% to 50% of Range
(40001)
IR – 5% to 95% of Range
(40001)
Range
2
GASES
USER MANUAL
Range
33