Priming Function - CME NIKI V4 Service Manual

Veterinary infusion pump
Table of Contents

Advertisement

Priming Function is used:
1. Before starting a program.
2. After air in line alarm. Disconnect IV tubing from patient, press "PRIME" key to
push air bubbles through.
Priming Function Operation:
1. Switch On the pump using ON/OFF key.
After a short self test the pump will display ON followed by
the last rate setting.
NOTE: Make sure the IV set is not connected to patient.
2. Press Prime.
3. The display will show FILL.
4. Once the priming is completed, the display will change
from FILL to the last rate setting.
5. Press the RUN/HOLD key to start operation.
NOTE: The pump will prime 5 ml instead of 16 ml in case the set is partially
Primed.
NIKI V4 Operator's Manual

Priming Function

24
RATE-
-