Figure 40 FA timer controls

The drop-down menu offers the following timer options:
• Start—Start the timer.
• Pause—Pause video capture, but keep the timer running.
• Restart—Restart the timer, following a stop or reset.
• Reset—Set the timer to zero again.
B. Move the fiber optic cable back to the white light port and switch OFF the FA
module power.
C. Remove and store the barrier filter in the special clear plastic box provided,
or in a manner that prevents damage. Clean the filter with compressed air
and optical cleaner and tissue as required.
When the Image Type selected is Fluorescein, the video and images are displayed
and stored in grayscale by default.
To view and store FA images and videos in color, select Fluorescein as the
Image Type instead of using the Switch to FA button.
WARNING: To mitigate the potential for excess light exposure, always start at the
lowest light intensity level and increase if necessary. Use only the shortest amount
of light necessary; no greater than 5 minutes.
WARNING: Do not look directly at the fiber optic port.
WARNING: If the light from the illumination source is not blue DO NOT USE IT
because there may be a filter failure.
WARNING: Do not attempt to override the interlock switches, or to operate in FA
mode with the cover off. The FA module contains internal sources of high voltage,
high current, heat, and moving parts. Only authorized technicians may attempt any
service or repairs to the FA module. Failure to observe this warning could result in
injury to operator or patient.
Fluorescein Angiography Module 55
PN 21-100375 Rev. A