54 Fluorescein Angiography Module
Lens piece
Barrier filter
Handpiece
+
+
+
RetCam3 User Manual

Figure 39 Barrier filter installation

Note: If lenses are exchanged during an FA imaging session, hold the camera
upright to avoid having the barrier filter fall out during the exchange.
3. Select Switch to FA and SNAP a "red-free" image (recommended).
4. Attach the fiber optic lamp connector from the handpiece into the FA module
fiber optic port. (See
5. Select Start Timer. Dye can now be injected.
Note: The timer can be paused or stopped at any time.
6. Adjust FA light intensity using the FA light intensity knob shown in
Note: You cannot adjust the FA light intensity using the footswitch or control
panel controls.
7. You can take still photos or video during the FA imaging session, and you can
SNAP still photos during video replay.
8. When the FA imaging session is completed:
A. Pause or Reset the timer.
Figure
38.)
Rubber O-ring
toward lens piece
Beveled edge
toward handpiece
Figure
38.