WARNING: Interference may occur in the vicinity of
equipment marked with the following symbol.
This equipment/system may cause radio interference or may disrupt the operation
of nearby equipment. It may be necessary to take mitigation measures such as:
• Reorient or relocate the receiving device
• Increase the separation between the equipment
• Connect the equipment into an outlet on a circuit different from that to which the
other device(s) are connected.
• Consult the manufacturer or field service technician for help.
An error message may be displayed, which then may require a system reset. To
perform a system reset, press the system power switch on the control panel for ~6
seconds until the system powers down, then press the switch again to power the
system back on.
WARNING: This equipment may cause radio interference or may disrupt the
operation of nearby equipment. It may be necessary to take mitigation measures,
such as reorienting or relocating the system or shielding the location.
WARNING: To avoid the risk of electric shock, this equipment must be connected to
a supply mains with protective earth.
WARNING: For continued protection against risk of fire, replace only with same
type and rating fuse.

Important User Safety Notices

1. Never contact the front of the lens piece with hard or sharp objects. This could
damage the precision optics and sealing.
2. DO NOT AUTOCLAVE any part of the device.
3. Appropriately power down and unplug the unit and allow the light source bulb to
cool before replacing it.
4. Always return the handpiece to the holster when not imaging. You can drape the
handpiece cable over the cord wrap post, but do not wrap it tightly or damage to
the fiber optic may result.
5. Do not block access to the holster.
6. The system is not intended to be transported outdoors.
7. Newly acquired data may be lost if power is interrupted during live imaging.
Note: backup battery power is available for at least 15 minutes after interrupt.