Gravity Acceleration Correction - AND SJ-3000WP/-BT Instruction Manual

Ip67 compact bench scale
Hide thumbs Also See for SJ-3000WP/-BT:
Table of Contents

Advertisement

11.2.Gravity Acceleration Correction

When the scale is first used or has been moved to another location, it should be calibrated using
a calibration weight. But if a calibration weight is not available, the gravity acceleration correction
will compensate the scale. Change the gravity acceleration value stored in the scale to the value
of the area where the scale will be used. Refer to the gravity acceleration map at the end of this
manual.
Note: Gravity acceleration correction is not required when the scale will be calibrated
using a calibration weight at the place where it is to be used.
Refer to "11.1.Calibration Mode" to enter the calibration
1.
mode. The CAL is displayed.
2.
Press the
TARE
value setting mode.
3.
Change the displayed value using the following switches.
To shift the blinking digit to the right.
MODE
To increase the value of the blinking digit by one.
SAMPLE
4.
Press the
TARE
returns to the newly stored gravity acceleration value.
5.
When calibration using a calibration weight is to be performed,
go to step 3 of "11.3.Calibration Using A Weight".
To finish the setting procedure, press and hold the
switch or the CAL switch. The scale returns to the weighing
mode.
switch to enter the gravity acceleration
switch. The display shows
and
30

Advertisement