Technical Specification - Pickering 65-218 User Manual

Lxi modular high voltage matrix
Table of Contents

Advertisement

pickering
LXI Modular High Voltage Matrix
y Modular Matrix Design With 4 Y Axis Connections
y User Configurable for X Dimensions by Plugin
Addition, up to 300 X Connections
y Capability to Use Internal Bus for Y
Connection Between Plugin Modules
y X & Y Loop-Thru Connections Allow Easy
Matrix Expansion
y Switch up to 2 A / 2 A Carry
y Cold Switch 750 VDC / Standoff 1000 VDC
for 1 s With Low Duty Cycle, When Not
Utilizing Chassis Internal Bus.
y Version With 1000 VDC Cold Switching /
1000 VDC Standoff
y Built-In Scan List Sequence Stores With
Triggering Capability
y Supported by BIRST ™ & eBIRST ™ Test Tools
y Fully Compliant to 1.5 LXI Standard
y 3 Year Warranty
The 65-218 is a modular platform that provides a large high voltage
matrix solution. It consists of a 65-200 scalable chassis which can be
populated with up to six 65-218 plugin matrix modules. The modules
are loaded into the front of the chassis giving the best flexibility and
ease of maintenance.
Each plugin module is a high voltage matrix with a carry current of
2 A plus higher pulse conditions - see specification table. The plugin
modules are 50x4 matrices and are available in the following types:
65-218-111 - no access to the chassis backplane and capable of cold
switching 750 VDC continuous. 65-218-101 - access to five separate
internal buses on the chassis backplane and limited to 400 VDC cold
switching. 65-218-121-CC - version with no backplane access and
1000 V cold switching.
The internal buses allow the Y connections of the matrix to be
interconnected to form larger matrix sizes. Six 65-218-101 plugin
modules in a 65-200 chassis can be linked to create a 300x4 matrix.
Y1
Y2
Y3
Y4
Y Isolation
Switches
Simplified Diagram For 65-218 Configured as a 300x4 Matrix
(Y Isolation Switches are only applicable to 65-218-101 Plugin)
SECTION 1 - TECHNICAL SPECIFICATION
X Isolation
Switches
X Bus Loop-Thru
LXI MODULAR HIGH VOLTAGE MATRIX 65-218
SECTION 1 - TECHNICAL SPECIFICATION
Pickering's Range of LXI High Voltage Matrices
Y-Axis
X Range
4
50 to 300
2
100, 200 or 300
4
75, 150 or 225
X and Y loop-thru connection on the plugin module's
front panel allow the linking of 50x4 matrices within
the same chassis or between adjacent chassis. This
allows easy matrix expansion in both the X and Y
directions.
Both types of plugin module have X isolation switches
between the matrix and the front panel connector. The
65-218-101 also has Y isolation switches for the front
panel connector and the six internal Y buses. Isolation
switches allow unused signal paths to be disconnected
to preserve signal integrity.
Configuring the Matrix
To select the parts that you need to create a matrix
simply:
• Specify a 65-200-002 LXI Scalable Chassis.
• Add Matrix Plugin Modules with or without chassis
internal Y-bus access as required.
Built-In Relay Self-Test - BIRST
The BIRST facility provides a quick and easy way of
finding relay failures. No test equipment is required,
simply unplug the front panel connectors, launch the
BIRST application and the tool will run a diagnostic
test that will find all relays with faulty contacts.
For more information go to:
Supported by eBIRST
This matrix is also supported by eBIRST. These tools
simplify switching system fault-finding by quickly
testing the system and graphically identifying the
faulty relay.
For more information go to:
65-218
Poles
Model No.
1
65-218
2
60-310
2
60-311
pickeringtest.com/birst
pickeringtest.com/ebirst
pickeringtest.com
Page 1.1

Advertisement

Table of Contents
loading