Pickering 65-218 User Manual

Lxi modular high voltage matrix

Advertisement

Quick Links

65-218 User Manual
LXI Modular High Voltage Matrix
pickeringtest.com
Issue 2.0 September 2023

Advertisement

Table of Contents
loading

Summary of Contents for Pickering 65-218

  • Page 1 65-218 User Manual LXI Modular High Voltage Matrix pickeringtest.com Issue 2.0 September 2023...
  • Page 2 © COPYRIGHT (2023) PICKERING INTERFACES. ALL RIGHTS RESERVED. No part of this publication may be reproduced, transmitted, transcribed, translated or stored in any form, or by any means without the written permission of Pickering Interfaces. Technical details contained within this publication are subject to change without notice.
  • Page 3 Pickering Interfaces strives to fulfil all relevant environmental laws and regulations and reduce wastes and releases to the environment. Pickering Interfaces aims to design and operate products in a way that protects the environment and the health and safety of its employees, customers and the public. Pickering Interfaces endeavours to develop and manufacture products that can be produced, distributed, used and recycled, or disposed of, in a safe and environmentally friendly manner.
  • Page 4 If this product is heavy reference should be made to the safety instructions for provisions of lifting and moving. STATIC SENSITIVE To indicate that static sensitive devices are present and handling precautions should be followed. REED RELAY MODULE 40-110/115/120/125 LXI MODULAR HIGH VOLTAGE MATRIX 65-218 Page iv Page (IV)
  • Page 5: Table Of Contents

    Section 6 Sequence Service & Triggering.........6.1 Section 7 Troubleshooting ..............7.1 LXI Status Indicators ..........7.1 LAN Reset ..............7.2 Section 8 Maintenance Information ...........8.1 Relay Look-up Tables ..........8.2 PCB Layout Diagram ..........8.4 BIRST Diagnostics ...........8.5 LXI MODULAR HIGH VOLTAGE MATRIX 65-218 Page v...
  • Page 6 THIS PAGE INTENTIONALLY BLANK LXI MODULAR HIGH VOLTAGE MATRIX 65-218 Page vi...
  • Page 7: Warnings & Cautions

    I/O signals that may be present. Blanking panels are available to order from Pickering. If the product is not used in this manner for example by using an extender card then additional care must be taken to avoid contact with exposed signals.
  • Page 8 AND USERS SHOULD CONNECT TO THE FRONT PANEL CONNECTORS WITH CONNECTORS OR CABLE ASSEMBLIES THAT ENSURE OPERATORS WILL NOT INADVERTENTLY COME IN CONTACT WITH HIGH VOLTAGES. LXI HIGH VOLTAGE 2-POLE MATRIX 60-310 Page viii LXI MODULAR HIGH VOLTAGE MATRIX 65-218 Page viii...
  • Page 9: Technical Specification

    3 Year Warranty X and Y loop-thru connection on the plugin module’s The 65-218 is a modular platform that provides a large high voltage front panel allow the linking of 50x4 matrices within matrix solution. It consists of a 65-200 scalable chassis which can be the same chassis or between adjacent chassis.
  • Page 10 Example Configuration 1 off 65-200-002 2U LXI Scalable Chassis 2 U Hex 50x4 Matrix with no chassis Y-bus access 6 off 65-218-111 Matrix Plugin, without chassis Y-bus (750 VDC or 1000 V Cold Switch) access (750 VDC cold switching) Hex Matrices that permit 4 concurrent connections between Y and X or 6 off 65-218-121-CC Matrix Plugin, without chassis or X and Y.
  • Page 11 2 U 300x4 Matrix with chassis Y-bus access (400 VDC Cold Switch) 1 off 65-200-002 2 U LXI Scalable Chassis Matrix that permit 4 concurrent connections between Y and X or X 6 off 65-218-101 Matrix Plugin, with chassis Y-bus and Y. Ability to expand matrix size using loop-thru connections access...
  • Page 12 Diagram showing the front panel cabling required to interconnect three 65-218-111 or 65-218-121-CC 50x4 matrix plugin modules as a single 150x4 matrix. The Y-bus Loop-Thru connections can be further expanded to other 65-218 Units. Overview of “Hot” & “Cold” Switching Techniques “Hot”...
  • Page 13 LXI Modular High Voltage Matrix, 65-218 Relay Type Matrix RF Specification - Single Plugin Module The 65-218 is fitted with high quality electro-mechanical Bandwidth (-3dB): (Measurement at front panel) relays. A spare relay is built onto the circuit boards to Plugin: 8.5 MHz...
  • Page 14 PXI switching and programmable resistor modules under LXI contol via a The 65-218 range is part of a growing family of Gigabit Ethernet interface. Also available are the 60-104, Scalable Matrix systems available from Pickering 60-105 &...
  • Page 15 SECTION 1 - TECHNICAL SPECIFICATION Appendix pickering LXI Modular High Voltage Matrix, 65-218 Sequencing Service and Triggering This unit features a software reset line which provides the ability for a user supplied reset to be applied to the The Sequencing service provides the user with the ability...
  • Page 16 To learn more, please go to: pickeringrelay.com Programming Pickering provide kernel, IVI and VISA (NI & Keysight) drivers which are compatible with all Microsoft supported versions of Windows and popular older versions. For a list of all supporting operating systems, please see: pickeringtest.com/os The VISA driver support is provided for LabVIEW Real Time Operating Systems (Pharlap and Linux-RT).
  • Page 17 Three Year Warranty & Guaranteed Long-Term Support All standard products manufactured by Pickering Interfaces are warranted against defective materials and workmanship for a period of three years from the date of delivery to the original purchaser. Extended warranty and service agreements are available for all our modules and systems with various levels to suit your requirements.
  • Page 18 SECTION 1 - TECHNICAL SPECIFICATION pickering THIS PAGE INTENTIONALLY BLANK Page 1.10 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 19: Functional Description

    SECTION 2 - TECHNICAL DESCRIPTION FUNCTIONAL DESCRIPTION The 65-218 is a modular switching system based on the 65-200 chassis and is designed to provide a high density flexible matrix solution by the addition of plugin modules. The component parts are as follows: •...
  • Page 20 Loop-Thru Connections Y Bus E Y1 to Y4 Front Panel Connections Y Isolation Relays Figure 2-1. 65-218-101 50x4 Matrix Plugin Module With Chassis Y Bus Access (Up to 400V Continuous Voltage Operation) Page 2.2 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 21 Loop-Thru Connections Y1 to Y4 Connections Figure 2-2. 65-218-111/65-218-121 50x4 Matrix Plugin Module Without Chassis Y Bus Access (Up to 750V Continuous Voltage Operation for 65-218-111 Up to 1kV Continuous Voltage Operation for 65-218-121-CC) Page 2.3 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 22 SECTION 2 - TECHNICAL DESCRIPTION pickering THIS PAGE INTENTIONALLY BLANK Page 2.4 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 23: Installation

    This can be downloaded via the web page: https://www.pickeringtest.com/product/65-200-002-lxi-2u-modular-chassis-6-slot HIGH VOLTAGE WARNING: When the 65-218 is connected to a test system, potentially leathal voltages may be present on any front panel connector - un-used connectors must be covered with the caps provided.
  • Page 24 SECTION 3 - INSTALLATION pickering THIS PAGE INTENTIONALLY BLANK Page 3.2 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 25: Programming Guide

    SECTION 4 - PROGRAMMING GUIDE PROGRAMMING OPTIONS FOR PICKERING INTERFACES LXI UNITS For information on the installation and use of drivers and the programming of Pickering’s products in various software environments, please refer to the Software User Manual. This is available as a download from: https://www.pickeringtest.com/support/software-drivers-and-downloads...
  • Page 26: Programming The 65-218

    PROGRAMMING THE 65-218 The 65-218 is a matrix unit which in its basic form has 50 connections on its X axis and 4 on its Y axis and a bank of 50 isolation switches on its X-axis. The 60-200 scalable chassis can be supplied with between one and six matrix...
  • Page 27 SECTION 4 - PROGRAMMING GUIDE pickering The model numbers and the corresponding programming sub-units used are summarised in the table below: Table 4-1. 65-218 Programming Sub-Unit Allocation 65-200 Scalable Chassis Populated With 65-218-111 or 65-218-121-CC Plugin Modules Card Sub-Unit Location in Bus/Device...
  • Page 28 Direct I/O Driver , piplx The Direct I/O driver is composed of a suite of separate libraries which interact to provide full control of a Pickering unit. In the case of a switch card the communications library picmlx, and the switching library piplx, are needed.
  • Page 29 Switch closures on the matrix are made using the pi40iv_Connect function and released using the pi40iv_ Disconnect function. The following code snippet demonstrates the minimum steps required to operate a switch on a Pickering unit. Firstly a device session is opened on the unit located at 192.168.1.100. Then a switch closure of the matrix point at (x,y) location (12,3) is made then released.
  • Page 30: Command Line Control

    COMMAND LINE CONTROL OF PICKERING LXI INSTRUMENTS For a description of command line control of a Pickering LXI device, please refer to the 65-200 Scalable Chassis user manual.
  • Page 31: Connections

    Y Bus & Y Loop-thru Pinout X Loop-thru Pinout (50-pin male D-type connector (9-pin male D-type connector (50-pin male D-type connector viewed from front of module) viewed from front of module) viewed from front of module) Page 5.1 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 32 SECTION 5 - CONNECTOR INFORMATION pickering THIS PAGE INTENTIONALLY BLANK Page 5.2 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 33: Sequence Service & Triggering

    For a description of the Sequencing Service and Triggering function as well information on the DIO/Trigger port, please refer to the 65-200 Scalable Chassis user manual. This can be downloaded via the web page: https://www.pickeringtest.com/kb/software-topics/pickering-sequence-manager Page 6.3 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 34 SECTION 6 - SEQUENCE SERVICE & TRIGGERING pickering THIS PAGE INTENTIONALLY BLANK Page 6.4 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 35: Troubleshooting

    Refer to the Warnings and Cautions at the front of this manual INSTALLATION PROBLEMS The Plug & Play functionality of Pickering switching products generally ensures trouble-free installation. If you do experience any installation problems you should first ensure that all cables are properly connected.
  • Page 36 SECTION 7 - TROUBLE SHOOTING pickering THIS PAGE INTENTIONALLY BLANK Page 7.2 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 37: Maintenance Information

    USING BIRST ™ (BUILT-IN RELAY SELF TEST) The Pickering Interfaces BIRST (Built-In Relay Self Test) Tool provides a simple means of testing selected Pickering Interfaces switching products, including this model. The tool can be used to find relay faults or to provide confirmation that a product is operating correctly, either as part of the regular system calibration or because a system fault is suspected and the switch system needs to be eliminated from the fault finding procedure.
  • Page 38 RL43 RL93 RL143 RL193 RL44 RL94 RL144 RL194 RL45 RL95 RL145 RL195 RL46 RL96 RL146 RL196 RL47 RL97 RL147 RL197 RL48 RL98 RL148 RL198 RL49 RL99 RL149 RL199 RL50 RL100 RL150 RL200 Page 8.2 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 39 X43 - Isolation RL243 X44 - Isolation RL244 X45 - Isolation RL245 X46 - Isolation RL246 X47 - Isolation RL247 X48 - Isolation RL248 X49 - Isolation RL249 X50 - Isolation RL250 Page 8.3 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 40 Y1 - Y4 Bus D Isolation RL270 RL269 RL268 RL267 Y1 - Y4 Bus E Isolation RL274 RL273 RL272 RL271 SPARE SPARE SPARE SPARE Figure 8.1 - High Voltage Matrix Plugin Module Component Layout Page 8.4 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...
  • Page 41: Birst Diagnostics

    External Loop-Back Connectors The 65-218 plugin module is fitted with the ability to perform self diagnosis of its switching matrix using the BIRST tool. This built-in diagnostic tool requires the user to remove all front panel cabling from the unit before running the application.
  • Page 42 SECTION 8 - MAINTENANCE INFORMATION pickering THIS PAGE INTENTIONALLY BLANK Page 8.6 LXI MODULAR HIGH VOLTAGE MATRIX 65-218...

Table of Contents