7.1 Reference conditions
7 Technical specifications
7.1
Reference conditions
Ambient tempera-
ture
Instrument status
7.2
Instrument
IC system
Material
Operating pres-
sure range
Intelligent com-
ponents
7.3
Ambient conditions
Operation
Ambient tem-
perature
Humidity
Storage
Ambient tem-
perature
Transport
Ambient tem-
perature
■■■■■■■■
88
The technical specifications listed in this chapter refer to the following ref-
erence conditions:
+25 °C (± 3 °C)
Operating > 40 minutes (in equilibrium)
Metal-free IC system
■
Compact system with modular design
■
Painted, CFC-free polyurethane hard foam, fire class V0
0 - 50 MPa (500 bar) high-pressure pump
■
0 - 35 MPa (350 bar) standard PEEK system
■
iPump, iDetector, iColumn, MagIC Net
+5 - +45 °C
20 - 80% relative humidity
–20 - +70 °C
–40 - +70 °C
■■■■■■■■■■■■■■■■■■■■■■
930 Compact IC Flex ChS (2.930.1200)
Need help?