3.17 Connecting the separation column
■■■■■■■■
44
A test chromatogram and a leaflet accompanies every column. Detailed
information on special IC applications can be found in the corresponding
"Application Bulletins" or "Application Notes". You can find these
online at
http://www.metrohm.com
them from your responsible Metrohm representative free of charge.
CAUTION
New separation columns are filled with a solution and sealed with stop-
pers on both sides. Before inserting the column, ensure that this solu-
tion can be mixed with the eluent being used (follow the information
provided by the manufacturer).
NOTE
Connect the separation column only after the initial start-up of the
instrument. Until that point, insert a coupling (6.2744.040) instead of
the guard column and separation column.
■■■■■■■■■■■■■■■■■■■■■■
in the Applications area or request
930 Compact IC Flex ChS (2.930.1200)