5.6 Servicing the high-pressure pump
Accessories
■■■■■■■■
60
3 Reinserting the inlet valve into the pump head
Insert the inlet valve into the inlet valve holder (the seal must be
visible) (1).
Screw the inlet valve holder up into the pump head (2) and
tighten it firmly by hand and then retighten it one additional ¼
turn using the adjustable wrench (3).
Refasten the connection capillary to the connection of the eluent
aspiration tubing (14-7).
Removing the pump head
For this step, you need the following accessories:
You can find these parts in the accessory kit: Vario/Flex Basic
(6.5000.000).
4 mm hex key (6.2621.030)
Removing the pump head
Prerequisites:
Is the high-pressure pump switched off?
Has the pressure been released?
Is the instrument switched off?
■■■■■■■■■■■■■■■■■■■■■■