3.10 Metrohm Suppressor Module (MSM)
3.10.2.4
Connecting the rinsing solution
■■■■■■■■
34
1 Connect the capillary labeled regenerant to the outlet of the inline fil-
ter (6.2821.120) using a pressure screw (6.2744.070).
2 Use a pressure screw (6.2744.070) to fasten a piece of the PTFE
capillary (6.1803.030) to the inlet of the inline filter.
3 Use a pressure screw (6.2744.070) and an adapter (6.2744.080) to
fasten the other end of the PTFE capillary to Port 1 of the Dosino.
4 Connect the FEP tubing (6.1805.120) from the bottle with the regen-
eration solution to Port 2 of the Dosino.
Various possibilities exist for rinsing the Metrohm Suppressor Module:
Rinsing solution via STREAM (recommended)
Use the eluent from the conductivity detector as rinsing solution.
Rinsing solution via Dosino
Prepare the rinsing solution in a separate bottle and convey it with a
Dosino.
The rinsing solution is connected to the rinsing solution capillary.
■■■■■■■■■■■■■■■■■■■■■■