Annual Maintenance; Pipetting Head Replacement / Calibration - Accuris AutoMATE 96 Operation Manual

Hide thumbs Also See for AutoMATE 96:
Table of Contents

Advertisement

4. Annual Maintenance

Contact the Accuris Instruments, or your authorized distributor or authorized service
provider, for any necessary maintenance, including but not limited to the following
items:
a. Check that the moving platform is level and operating smoothly.
b. Check the pistons and seals to make sure there is no leakage, and that pipetting
volumes are consistent between channels.
c. Calibrate the instrument as required to make sure it meets specifications.
5. Pipetting Head Replacement & Calibration
In the Settings mode, select "Maintenance" to enter the maintenance interface.
Base calibration and Pipettor calibration adjustments should be performed by
authorized service personnel. Contact Accuris Technical Services or a local
distributor for more information.
To switch a pipetting head, click Pipettor calibration and click the button "Replace
module". Follow the flow chart for pipettor head replacement in Section 2. Click the
icon
to scan the QR code label on the pipetting module using the rear camera
of the tablet. This automatically loads the specific calibration data for the pipetting
module.
Fig 45 Pipettor calibration
Fig 46 Prompt box for pipette head replacement
37