• Bottom (mm): Adjust how far the tips are placed from the bottom of the
labware used when performing aspiration.
Dispensation Step Parameters:
The following settings are available and adjustable for the dispensation steps:
• Position: Platform position (P1 – P4)
• Labware List: Select from which plate dispensation will occur (96 deep well
plate, reagent reservoir, etc.)
• Dispense Volume (µL): The set volume for dispensation
• Blow Liquid Volume (µL): The set volume for blowing out residual liquid in
the pipette tips after performing dispensation
• Dispense Position: Select where dispensing will occur: "Top" or "Bottom"
• Top (mm): Adjust How far from the top of the dispense position selected the
AutoMATE 96 will dispense
• Bottom (mm): Adjust How far from the bottom of the dispense position
selected the AutoMATE 96 will dispense
• Post-Mix: Toggle to enable mixing of the reagent post-dispensation
o Mix Volume (µL): The volume of the reagent that will be mixed after
dispensation
o Cycles: The # of post-mix cycles performed after completing
Figure 20 Dispense Settings