3-26 Testing the HP 8133A Pulse Generator
Figure 3-21.
BIT RZ
32
1. Set up the HP 8133A Pulse Generator
2. Set Bits 0 to 30 of the 32 Bit data to '1' and Bit 31 to 0:
3. On the oscilloscope:
a. Press @ - ) .
b. Select the Timebase menu and set the TIME/DIV to 500 ps/div
and the DELAY to 16 ns.
c. Select th.e Delta
V
d. Set the Preset Levels to 50-50% and press Auto Level Set.
e. Select thle Delta t menu and switch on the T Markers.
HP 5 4 1 2 1 T
O s c i 1
2 0 d B A t t e n u a t o r s

Channel Data Test Set-up

as follows:
SQUAR
FREQ 3.000 GHz
HIGH +0.6
V
LOW -0.5 V
BIT
0
Off (Output enabled)
DATA
HIGH
V
V
32 BIT
Off (OUTPUT enabled:
menu and switch on the V Markers.