Data Performance

r. Set ithe Window Marker 2
of the signal amplitude around
s. Select Acquire and set the Number of Samples to 1000
t. Press Start Acquiring.
u. After 1000 samples are completed, select Results and press
Sigma:
Figure 3-20. Jitter Histogram
89. Record the Sigma value on Test Record as the RMS Jitter for the
Sigma is not. actually the true RMS Jitter value.
The Sigma value includes the effect
window, and the oscilloscope's triggerkimebase jitter.
If the Sigma, value is not within specification, make a precise
calculation of the true RMS Jitter:
a. Measure and eliminate the falltime error:
-
(6
x
6
b. Eliminate the oscilloscope's Jitter:
Timebase Jitter
=
Refer to Product Note 54120-2 for information on how to measure
the Timebase
and
Trigger Jitter of the oscilloscope.
c. Indicate on the Test Record that you have corrected the measured
Sigma to the true RMS Jitter value.
The Data Performance test applies only to
(Option 002
to -15
0
V.
5 ps.
the falltime of the f 1 5 mV
of
-
HP 8133A Pulse
Testing the
Channel 2
Generator
3-25