Chapter 4. Programming

4.6 Programming the PLACEMENT Menu

The PLACEMENT menu allows the user to configure the mounting method of the
transducers, based on the programmed information in the TRANSDUCERS menu
(see "Programming the TRANSDUCERS Menu" on page 66).

4.6.1 Viewing the Traverse Configuration

For CLAMP-ON transducers, one of the six possible TRAVERSE configurations
shown in Figure 79 below is displayed, as appropriate for your programmed
transducer information. Typically, a two-traverse installation is used.
Figure 79: Clamp-On Traverse Configurations
For WETTED transducers, the values for the following parameters are listed, as
calculated from your programmed transducer information:
PATH LENGTH
AXIAL LENGTH
72
®
TransPort
PT900 User's Manual