Chapter 8. Error Codes and Troubleshooting

8.1 Error Codes

The bottom line of the Tablet LCD displays a single, top-priority error message
during operation. This line, called the Error Line and it includes the following:

Error Header:

Error String:
8.1.1 Error Header
The possible Error Headers are listed in Table 15 below.

8.1.2 Flow Errors

Flow Errors are errors which occur while taking flow measurements. These errors
can be caused by disturbances in the fluid, such as excessive particles in the flow
stream or extreme temperature gradients. The errors could also be caused by an
empty pipe or issues with the fluid. Flow errors are typically not caused by a
malfunction of the flow measurement device, but by an issue with the fluid itself.
8.1.2a
E1: Low Signal
Problem:
Poor ultrasonic signal strength or the signal exceeds the programmed
limit.
Cause:
SNR is less than the Signal Low Limit or the signal cannot be found.
Poor signal strength may be caused by a defective cable, a fluid or pipe
problem, a defective transducer or a transmitter problem. A signal that
exceeds the programmed limits is probably caused by an improper user
value.
Action:
Check the components listed above. Also, check the programmed error
limits value.
®
TransPort
PT900 User's Manual
gives the Error Pattern and the Error Number
gives detailed error information
Table 15: Error Headers