10.1.5
Receiving data from SEMA
Patient data can be received from a Schiller SEMA server and automatically entered
on the CARDIOVIT FT-1. This is called patient data query (PDQ). To do this, the
Patient or Visit ID is entered in the patient data screen manually or via a barcode
reader (see
Page
• Patient data query requires the Schiller SEMA Server being available via network.
• The server name, URL, TCP/IP address etc. as well as all other transmission set-
tings are defined in the system settings (see
• A communication overview is given in the SCHILLER Communication Handbook
(Art. No. 2.520036).
10.1.6
Failed data transmission
Failed transmission of data is indicated with the symbol
1. The blue figure indicates the number of failed transmissions.
2. Data that was not transmitted can be resent from the memory. See
Transmitting and deleting stored recordings, page 67.
If no data can be transmitted, check the following:
Network settings (see
Network connection WLAN or LAN
Encryption settings on the server
Settings in the Schiller Link App.
28).
Page 82)
Transmission - Overview
Transmission Options
Page
82).
in the status bar.