Resting ECG

4.11 Lead sequence/lead view
Page 40
5 Resting ECG
 After heavy artefacts or lead off, the displayed heart rate may not be reliable.
 The safety notes at the beginning of this book must be read and fully understood
before taking an ECG recording.
 The CARDIOVIT FT-1 device is CF classified
fully isolated. During the ECG recording, ensure that neither the patient nor the
leading parts of the patient connection nor the electrodes (including the neutral
electrodes) come in contact with other persons or conductive objects, even when
these are earthed.
 Do not operate the unit if the earth connection is suspect or if the mains lead is
damaged or suspected of being damaged.
 If an external electronic device (e.g. a PC) is connected to the CARDIOVIT FT-1,
use the potential equalisation stud for earth protection.
If another format than the default format is set for the automatic printout, the printout
can differ from the format displayed on the screen.
The displayed lead sequence (Standard or Cabrera) can be selected (see
Leads & cable Page
Views & layouts, Page
For the preview, the following parameters can be freely programmed (before start of
the recording):
• Lead configuration
• Filter
Saved recordings can be displayed and printed in another format at any time. For
further information on how to define the format, see
 When using the 25 Hz or 40 Hz filter, the displayed or printed ECG may not
always fulfil the requirements for a diagnostic ECG.
51). Amplitude and speed can be defined in the menu
53.
CARDIOVIT FT-1
. The patient connection is
5.6.7ReportsPage
57.
5.6.1
5.6.3