Fluke 5560A Service Manual page 75

Hide thumbs Also See for 5560A:
Table of Contents

Advertisement

Table 31. Calibration Adjust Steps for Capacitance with LCR meter
Step
1
0.2 nF
2
1 nF
3
1.21 nF
4
10 nF
5
12.1 nF
6
20 nF
7
30 nF
8
50 nF
9
100 nF
10
120 nF
11
0.121 µF
12
1 µF
13
1.21 µF
14
10 µF
15
12.1 µF
16
100 µF
17
0.121 mF
18
1 mF
19
1.21 mF
20
10 mF
21
12.1 mF
22
100 mF
Product Output
Calibration Adjustment
Test Frequency
1000 Hz
1000 Hz
1000 Hz
1000 Hz
610 Hz
610 Hz
610 Hz
610 Hz
610 Hz
610 Hz
100 Hz
100 Hz
80 Hz
80 Hz
20 Hz
20 Hz
5 Hz
5 Hz
2 Hz
2 Hz
1 Hz
1 Hz
Calibrator