Fluke 5560A Service Manual page 55

Hide thumbs Also See for 5560A:
Table of Contents

Advertisement

Product
Product
Step
Range
Output
34
12 mA
10 mA
35
12 mA
10 mA
36
12 mA
10 mA
37
12 mA
10 mA
32
120 mA
12.1 mA
38
120 mA
12.1 mA
39
120 mA
12.1 mA
40
120 mA
12.1 mA
41
120 mA
12.1 mA
42
120 mA
12.1 mA
43
120 mA
100 mA
45
120 mA
100 mA
46
120 mA
100 mA
47
120 mA
100 mA
48
120 mA
100 mA
49
120 mA
100 mA
Table 16. Calibration Steps for Amps AC (cont.)
Product
Shunt
Frequency
Fluke A40B-
1 kHz
10mA
Fluke A40B-
5 kHz
10mA
Fluke A40B-
10 kHz
10mA
Fluke A40B-
30 kHz
10mA
Fluke A40B-
3 Hz
10mA
Fluke A40B-
45 Hz
10mA
Fluke A40B-
1 kHz
10mA
Fluke A40B-
5 kHz
10mA
Fluke A40B-
10 kHz
10mA
Fluke A40B-
30 kHz
10mA
Fluke A40B-
3 Hz
100mA
Fluke A40B-
45 Hz
100mA
Fluke A40B-
1 kHz
100mA
Fluke A40B-
5 kHz
100mA
Fluke A40B-
10 kHz
100mA
Fluke A40B-
30 kHz
100mA
Calibration Verification of the Product
DMM
Calculated
Shunt_R
Reading on
[Ω]
Shunt
Output [V]
Calibrator
Product
Limits
Output
±3.00 μA
±3.00 μA
±13.00 μA
±50 μA
±12 μA
±6 μA
±6 μA
±10 μA
±25 μA
±148 μA
±30 μA
±17 μA
±17 μA
±28 μA
±130 μA
±500 μA
51