24. With the entire table populated, evaluate each step for compliance to limits. If Guardbanding is to
be applied create a table from the example in this manual, where the acceptance limits are properly
modified by your Guardbanding method.
Product
Step
Range
1
120 mV
0 mV
2
120 mV
-0.00001 mV
3
120 mV
10 mV
4
120 mV
-10 mV
25
120 mV
100 mV
26
120 mV
-100 mV
5
1.2 V
0 V
6
1.2 V
-0.0000001 V N/A
27
1.2 V
0.1 V
28
1.2 V
-0.1 V
29
1.2 V
1 V
30
1.2 V
-1 V
7
12 V
0 V
8
12 V
-0.000001 V
31
12 V
1 V
32
12 V
-1 V
33
12 V
10 V
34
12 V
-10 V
9
120 V
0 V
10
120 V
-0.00001 V
35
120 V
10 V
36
120 V
-10 V
11
120 V
100 V
12
120 V
-100 V
13
120 V
120 V
14
120 V
-120 V
15
1020 V
0 V
16
1020 V
-0.0001 V
14
Table 3. Calibration Steps for Volts DC (OUTPUT VZ)
Product
732C Value
Output
(Vstd)
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
DMM Reading
DMM Reading
on 732C
on Product
(Vdmm_732C)
(Vdmm_prod)
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
Calculated
Product
Limits
Output
±0.80 µV
±0.80 µV
±0.90 µV
±0.90 µV
±1.8 µV
±1.8 µV
±1.0 µV
±1.0 µV
±1.7 µV
±1.7 µV
±7.6 µV
±7.6 µV
±10.0 µV
±10.0 µV
±17 µV
±17 µV
±75 µV
±75 µV
±0.10 mV
±0.10 mV
±0.19 mV
±0.19 mV
±1.03 mV
±1.03 mV
±1.22 mV
±1.22 mV
±1.00 mV
±1.00 mV