5560A/5550A/5540A
Service Manual
5560A
22. Set the Product to Standby. Disconnect the shunt. Connect the test equipment as shown in
Figure 9.
23. With no current applied from the Product to the Fluke A40B-100mA shunt, proceed with the zero
meter function. On the 8588A, push ZERO, and then ZERO FUNC. Allow the zero function to finish.
24. Start with step 31. Set the output from the DUT/Measure with the meter and record the 8588A
measured value in the DMM Reading on Shunt Output [V] column. Complete all steps for the
present connection (31 to 34). Refer to the Product manual for how to set/use range lock on bottom
of range outputs.
5560A
20
Figure 8. Amps DC with A40B-10mA
A40B-10mA
A40B-10mA
Figure 9. Amps DC with A40B-100mA
A40B-100mA
8588A
8588A