Amps Ac Calibration - Fluke 5560A Service Manual

Hide thumbs Also See for 5560A:
Table of Contents

Advertisement

Table 14. Calibration Steps for Volts AC Out of the AUX Output (cont.)
Product Output
Step
(OUTPUT VZ)
9
1 V
10
1 V
11
1 V
12
1 V
13
1 V
14
1 V

Amps AC Calibration

Table
15
lists the required equipment.
Equipment Description
Digital Multimeter
(1 mA to 1.2 mA)
(10 mA to 12 mA)
(100 mA to 120 mA)
(1 A to 1.2 A)
3.1 A, 3.11 A
10 A
(12.1 A to 20 A)
30 A
Type N (M) to dual banana (M)
Type N (F) to Type N (F) adapter
LC Male to LC Male adapter
LC Female to N Male inter-series adapter
N to 4 mm double banana connector
DBL BANANA N (M) RG58C/U (≈ 16 in)
The amps ac calibration uses two methods. For frequencies ≤ 10 Hz, use a low-frequency Fluke
Calibration 8588A meter to monitor the output of Fluke A40B series shunts. For frequencies >10 Hz the
calibration is accomplished with the Fluke Calibration 5790B AC Measurement Standard monitoring
the output of the Fluke A40B series shunts.
Product Output
(AUX Output)
5 V
45 Hz
5 V
400 Hz
5 V
1 kHz
5 V
5 kHz
5 V
10 kHz
5 V
30 kHz
Table 15. Required Equipment for Amps AC
Product
Product Output
Frequency
Manufacture/Model/PN
Fluke 8588A
Fluke A40B-1mA Current Shunt
Fluke A40B-10mA Current Shunt
Fluke A40B-100mA Current Shunt
Fluke A40B-1A Current Shunt
Fluke A40B-5A Current Shunt
Fluke A40B-10A Current Shunt
Fluke A40B-20A Current Shunt
Fluke A40B-50A Current Shunt
E-Z-Hook PN 9415
Various
A40B-ADAPT/LC
A40B-ADAPT/LCN
A40B-LEAD/4mm
(Fluke PN 900394) 5790-8026
Calibration Verification of the Product
±4.05 mV
±4.05 mV
±4.05 mV
±7.9 mV
±15.4 mV
±188 mV
Calibrator
Limits
Quantity
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
41