Fluke 5560A Service Manual page 29

Hide thumbs Also See for 5560A:
Table of Contents

Advertisement

Product
Product
Step
Range
Output
9
1.2 mA
0.1 mA
10
1.2 mA
-0.1 mA
23
1.2 mA
1 mA
24
1.2 mA
-1 mA
11
12 mA
0 mA
12
12 mA
-0.00001 mA
25
12 mA
1 mA
26
12 mA
-1 mA
27
12 mA
10 mA
28
12 mA
-10 mA
13
120 mA
0 mA
14
120 mA
-0.0001 mA
29
120 mA
10 mA
30
120 mA
-10 mA
31
120 mA
100 mA
32
120 mA
-100 mA
15
1.2 A
0 A
16
1.2 A
-0.000001 A
33
1.2 A
0.1 A
34
1.2 A
-0.1 A
35
1.2 A
1 A
36
1.2 A
-1 A
17
3.1 A
0 A
18
3.1 A
-0.00001 A
37
3.1 A
1 A
38
3.1 A
-1 A
Table 7. Calibration Steps for Amps DC (cont.)
Shunt
Fluke 742A-1k
Fluke 742A-1k
Fluke A40B-1mA
Fluke A40B-1mA
Fluke 742A-1k
Fluke 742A-1k
Fluke A40B-1mA
Fluke A40B-1mA
Fluke A40B-10mA
Fluke A40B-10mA
Fluke 742A-1k
Fluke 742A-1k
Fluke A40B-10mA
Fluke A40B-10mA
Fluke A40B-100mA
Fluke A40B-100mA
Fluke 742A-1k
Fluke 742A-1k
Fluke A40B-100mA
Fluke A40B-100mA
Fluke A40B-1A
Fluke A40B-1A
Fluke 742A-1k
Fluke 742A-1k
Fluke A40B-1A
Fluke A40B-1A
Calibration Verification of the Product
DMM
Reading
Calculated
Shunt_R
on Shunt
[Ω]
Output
[V]
Calibrator
Product
Limits
Output
±23.0 nA
±23.0 nA
±95.0 nA
±95.0 nA
±0.08 µA
±0.08 µA
±0.16 µA
±0.16 µA
±0.88 µA
±0.88 µA
±0.80 µA
±0.80 µA
±2 µA
±2 µA
±9 µA
±9 µA
±10 µA
±10 µA
±23 µA
±23 µA
±140 µA
±140 µA
±150 µA
±150 µA
±390 µA
±390 µA