Fluke 5560A Service Manual page 27

Hide thumbs Also See for 5560A:
Table of Contents

Advertisement

38. Set the Product to Standby. Disconnect the shunt. Connect the test equipment as shown in
Figure 13.
39. With no current applied from the Product to the Fluke A40B-20A shunt, proceed with the meter zero
function. On the 8588A, push ZERO, and then ZERO FUNC. Allow the zero function to finish.
40. Start with step 45. Set the output from the DUT and wait 3 minutes for settling. Measure with the
meter and record the 8588A measured value in the DMM Reading on Shunt Output [V] column.
Complete the steps for the present connection (45 to 50). Refer to the Product manual how to set
and use range lock on bottom of range outputs.
5560A
41. Set the Product to Standby. Disconnect the shunt. Connect the test equipment as shown in
Figure 14.
42. With no current applied from the Product to the Fluke A40B-50A shunt, proceed with the meter zero
function. On the 8588A, push ZERO, and then ZERO FUNC. Allow the zero function to finish.
43. Start with step 51. Set the output from the DUT and wait 3 minutes for settling. Measure with the
meter and record the 8588A measured in the DMM Reading on Shunt Output [V] column. Complete
both steps for the present connection (51 and 52). Refer to the Product manual how to set/use
range lock on bottom of range outputs.
Figure 13. Amps DC with A40B-20A
A40B-20A
A40B-20A
INPUT
OUTPUT
OUTPUT
INPUT
Calibration Verification of the Product
8588A
Calibrator
23