Fluke 5560A Service Manual page 54

Hide thumbs Also See for 5560A:
Table of Contents

Advertisement

5560A/5550A/5540A
Service Manual
Product
Product
Step
Range
Output
19
1.2 mA
0.121 mA
20
1.2 mA
0.121 mA
4
1.2 mA
1 mA
21
1.2 mA
1 mA
22
1.2 mA
1 mA
23
1.2 mA
1 mA
24
1.2 mA
1 mA
25
1.2 mA
1 mA
5
12 mA
1.21 mA
26
12 mA
1.21 mA
27
12 mA
1.21 mA
28
12 mA
1.21 mA
29
12 mA
1.21 mA
30
12 mA
1.21 mA
31
12 mA
10 mA
33
12 mA
10 mA
50
Table 16. Calibration Steps for Amps AC (cont.)
Product
Shunt
Frequency
Fluke A40B-
10 kHz
1mA
Fluke A40B-
30 kHz
1mA
Fluke A40B-
3 Hz
1mA
Fluke A40B-
45 Hz
1mA
Fluke A40B-
1 kHz
1mA
Fluke A40B-
5 kHz
1mA
Fluke A40B-
10 kHz
1mA
Fluke A40B-
30 kHz
1mA
Fluke A40B-
3 Hz
1mA
Fluke A40B-
45 Hz
1mA
Fluke A40B-
1 kHz
1mA
Fluke A40B-
5 kHz
1mA
Fluke A40B-
10 kHz
1mA
Fluke A40B-
30 kHz
1mA
Fluke A40B-
3 Hz
10mA
Fluke A40B-
45 Hz
10mA
DMM
Calculated
Shunt_R
Reading on
[Ω]
Shunt
Output [V]
Product
Limits
Output
±245 nA
±5484 nA
±300 nA
±300 nA
±300 nA
±300 nA
±1300 nA
±9.00 μA
±1.24 μA
±1.24 μA
±1.24 μA
±1.24 μA
±2.45 μA
±14.84 μA
±3.00 μA
±3.00 μA

Advertisement

Table of Contents
loading

This manual is also suitable for:

5550a5540a

Table of Contents