Fluke 5560A Service Manual page 31

Hide thumbs Also See for 5560A:
Table of Contents

Advertisement

Table
8
lists the required equipment.
Equipment Description
Digital Multimeter
Resistance Standard
Resistance Standard
Resistance Standard
Resistance Standard
Resistance Standard
Low Thermal Cables
Double banana RG58 cable
BNC (F) to Type N (M)
BNC (F) To Single Banana Plug
Resistance Standard.
The instructions in this section and the measurement sequence is structured to use common settings,
modes, and methods to simplify the calibration function.
To calibrate the Resistance function:
1. Make sure that the DUT is in Standby.
2. Set the 8588A as follows: Resistance 4Wire, LoI OFF, Range (as per Table 10), NPLC 50, Guard
OFF (external guard light ring off). Set the meter to DC Voltage temporarily to prevent compliance
limit errors on the Product.
3. Connect a strap between the Guard and Ground terminals of the DUT.
4. Connect the test equipment as shown in Figure 15.
5. Start with step 1 in Table 10. Set the output from the DUT 4W Resistance. Measure with the meter
and record the 8588A measured value in the DMM Measurement on Product column in Table 10.
Change the meter range for each step. Complete all steps for the present connection and meter
setup (1 through 14).
Table 8. Required Equipment for Resistance
Fluke Calibration 8588A
Fluke 742A-100k
Fluke 742A-1M
Fluke 742A-10M
MI 9331G/100 M
MI 9331G/1 G
Fluke Calibration 5730A-7003
Pomona Model 2BC-24 and 2BC-36
Pasternack PN PE9002
Pomona Model 1894
Fluke 742-10k
Calibration Verification of the Product
Manufacture/Model/PN
Calibrator
Quantity
1
1
1
1
1
1
1
1
2
1
1
27