7. Be sure to connect the power plug to the dedicated power distribution panel or
a wall outlet.
Use a power distribution panel or a wall outlet that meets the electrical capacity of the unit.
Electrical
capacity:
* When the unit will not start even when you turn the ground fault interrupter to "ON",
check for low main voltage or if the unit is connected to the same power supply line
as other devices and connect it to another line if necessary.
Avoid connecting too many devices using a branching outlet or extending a wire with a cord reel
or temperature controlling function may degrade due to voltage drop.
Do not connect the unit to a gas pipe, a water pipe or a telephone line or any parts or lines,
other than a correct power supply line.
Otherwise, an accident or a malfunction may result.
8. Handling of a power cord
Never use electrical power cords bundled. When these are used bundled, they might
overheat causing a fire.
Do not convert, forcibly bend, twist or pull the power cord. Otherwise, a fire or an electrical
shock may result.
Do not place the power cord under a desk or a chair, or sandwich between objects to avoid
it from being damaged. Otherwise, a fire or an electrical shock may result.
Do not place the power cord close to a stove or other heat generating device. Sheath of
the cord may burn and result in a fire or an electrical shock.
If the power cord should be damaged (exposure of core wire or disconnection), immediately
turn the power switch off, pull out the power cord (plug) out of the power supply and ask
your dealer to replace the cord. Otherwise, a fire or an electrical shock may result.
Connect the power cord to an appropriate wall outlet.
9. Be sure to connect the ground wire.
・ When the unit has no ground terminal, class D grounding works is necessary and
please consult your dealer or our nearest sales office.
・ Securely connect to an outlet.
We recommend use of a ground type outlet
tap.
When there is no ground terminal.
In this case, class D grounding works is
necessary and please consult your dealer or
our nearest sales office.
Do not connect the grounding wire to a gas pipe, a water pipe or a telephone line or any
parts or lines other than a correct grounding terminal.
Otherwise, an accident or a malfunction may result.
2. Before operating the unit
IC103C/103CW
AC115V 1.8A
IC403C/403CW
AC115V 4.5A
IC603C
AC115V 6A
IC803C
AC115V 10A
Grounded
tap
Power
plug
Precautions when installing the unit
IC113C/113CW AC220V 1A
IC413C/413CW AC220V 1.4A
IC613C
IC813C
When a bipolar type outlet tap is used
Bipolar
tap
Insert the ground adaptor included as an
option, into a power plug confirming the
polarity of the outlet. Connect the grounding
wire (green) of the ground adaptor to the
ground terminal on the power supply
equipment.
6
AC220V 1.8A
AC220V 3.3A
Ground wire

Advertisement