1. About handling of flammable or combustible solution
The unit is not explosion proof. Take special care for handling samples on which explosive
substances, combustible substances or substances containing them.
combustible solution will evaporate when left at a room temperature (or at a lower
temperature for some types of solutions) and may be ignited and explode from switches,
lights and other ignitable sources. Be sure to assure sufficient ventilation when using
these materials.
See section "13. List of dangerous materials" on page 41.
2. Ban on use/countermeasures when an error occurs
If smoke emerges from the unit or an odd odor is felt, immediately turn the power switch on
the main unit off, turn the power supply off and contact your dealer or a Yamato sales office
for inspection. Otherwise, a fire or an electrical shock may result. The user shall never
attempt to repair the unit to avoid any possible dangers.
3. Secure sufficient ventilation for the unit.
Do not operate the unit when its vent holes on the side and rear panels covered or blocked.
Internal temperature of the unit will rise degrading the performance and an accident, a
malfunction or a fire may result.
4. Do not allow liquid to spill over the unit.
Do not allow liquid to spill over the unit. Pay special attention not to allow liquid to enter
into the vent holes on the side and rear panels of the unit. If liquid is spilt over or into the
unit, do not try to operate it any further. Other wise, an accident, a malfunction, a fire or an
electrical shock may result.
5. Do not allow a metal piece to fall into the unit.
Do not allow a clip, a staple, a screw or other metal pieces to fall into the unit.
Stop operating the unit if a metal piece has dropped into the unit.
Other wise, an accident, a malfunction, a fire or an electrical shock may result.
6. Do not open the cabinet.
Do not open panels or covers fixed on the unit, or do not operate the unit with any of those
open. Other wise, an accident, a malfunction, or an electrical shock may result.
7. Always operate the unit at a correct ambient temperature.
The operating temperature range is room temperature range from +5 ~ 80 ℃ above room
temperature. Never try to operate the unit outside the operating temperature range.
8. Do not attempt to modify the unit.
The user shall never try to modify the unit; other wise, an accident, a malfunction, a fire or
an electrical shock may result.

5. Cautions on handling