100
12
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation in the table above applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
USM-100H is intended for use in the electromagnetic environment specified below. The customer or
the user of the model USM-100H should assure that it is used in such an environment.
Immunity
IEC 60601 Test Level
Test
Electrostatic
dis-charge
±8 kV contact
(ESD)
±15 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power supply lines
transient/burst
±1 kV for input/output lines
IEC 61000-4-4
Surge
±1 kV line(s) to line(s)
IEC 61000-4-5
±2 kV line(s) to earth
Voltage Dips:
i) 0% reduction for 0.5 cycle
at 50Hz,
0% reduction for 1 cycle
Interruptions
at 50Hz
and voltage
Performance Criterion B
variations on
ii) 70% reduction for 25/30
power supply
cycles at 50/60Hz
input lines
Performance Criterion C
IEC 61000-4-11
Voltage Interruptions:
0% reduction for 250/300
cycles at 50/60Hz
Performance Criterion C
Power
frequency
(50/60 Hz)
30A/m
magnetic field
IEC 61000-4-8
NOTE UT is the A.C. main voltage prior to application of the test level.
12
23
Compliance Level
±8 kV contact
±15 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Voltage Dips:
i) 0% reduction for 0.5 cycle
at 50Hz,
0% reduction for 1 cycle
at 50Hz
Performance Criterion B
ii) 70% reduction for 25/30
cycles at 50/60Hz
Performance Criterion C
Voltage Interruptions:
0% reduction for 250/300
cycles at 50/60Hz
Performance Criterion C
30A/m
17
Electromagnetic
Environmental
Guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Main power quality should
be that of a typical
commercial or hospital
environment.
Main power quality should
be that of a typical
commercial or hospital
environment.
Main power quality should
be that of a typical
commercial or hospital
environment. If the user of
the model USM-100
requires continued
operation during main
power interruption, it is
recommended that the
model USM-100 be
powered from an
uninterruptible power
supply.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
POC-421_424 User Manual