Automatic Calibration/Adjustment With Isocal - Sartorius SECURA1102-1NUS User Manual

Laboratory balances
Table of Contents

Advertisement

Calibration and Adjustment
Max 2100 g
isoCAL
0.00
Report
Date and Time:
14-Jun-2012
Calibration function:
Internal calibration
Start calibration:
Start: isoCAL (Niv)
Calibration deviation:
Dev -0.01 g
Adjustment function:
Internal adjustment
Adjustment deviation:
Dev 0.00 g
76
User Manual Secura
®
, Quintix

8.3 Automatic Calibration/Adjustment with isoCAL

Fully automatic calibration/adjustment ensures that the balance is automatically
calibrated and adjusted when predefined time intervals or temperature values are
exceeded .
This function varies in different models:
Secura
/ Quintix
®
temperature, and each time the balance is leveled . This function can be switched
off via the system settings (see page 31) .
In Secura
®
models, every adjustment process, including the identified deviations,
is documented and saved on the balance . The saved reports can be viewed via the
system settings and printed out on a PC or Sartorius lab printer (see page 31) .
y If the ambient temperature has changed since the last calibration/adjustment or
the preset time interval has been exceeded, the red
d = 0.01 g
the display .
isoCAL
y If the
starts automatically .
! If the balance has not been adjusted, the prompt will be shown again 5 minutes
after the "isoCAL" status message is acknowledged .
y If the
isoCAL
manually .
To start the calibration/adjustment function manually, select the
isoCAL
t
(Secura
®
, Quintix
y Calibration starts automatically .
y A report appears on the display (Secura
The report indicates the deviations identified during isoCAL calibration .
15:18
The report also gives the result of the adjustment process .
Close the report window: Select
t
y The balance is now calibrated and adjusted .
®
, Practum
®
: isoCAL is automatically carried out based on time or
®
function is set to Automatic, the calibration/adjustment process
function is set to
Info, manual start
®
) .
and Quintix
®
<
.
isoCAL
button appears on
, the function must be started
isoCAL