Leveling; Leveling With Electronic Level Indicator (Secura ® ) - Sartorius SECURA1102-1NUS User Manual

Laboratory balances
Table of Contents

Advertisement

LEVEL
Max 2100 g
d = 0.01 g
0.00
LEVEL

Leveling

Please turn each foot in the
direction of the arrow as
shown.
Leveling
The balance has been leveled.
5.2 Leveling
The balance must be level to ensure precise weighing results . Leveling the balance
compensates for slant or unevenness at the place of installation by twisting the front
leveling feet of the balance .
The balance needs to be re-leveled and adjusted each time its setup location is
changed .
The leveling function is different for the various balance models:
Electronic level indicator (Secura
Conventional level indicator (Quintix
Note: The balance needs to be re-leveled and then adjusted each time its location is
changed .
5.2.1
Leveling with Electronic Level Indicator (Secura
In Secura
models, the balance's position is checked with sensors . Instructions on the
®
display guide the user when leveling the balance .
LEVEL
y When
appears in red at the top left of the display, you must level the
balance .
To start the leveling function, select the
t
y The balance position is shown as an animation of a level indicator on the display .
Twist the front leveling feet of the balance in the specified direction .
t
A prompt appears every 2 minutes after confirmation of this "Leveling" status
message if the balance has not been leveled .
y When the level indicator animation turns green, the balance is correctly leveled .
To exit the function, select
t
Complete the leveling by calibrating the balance .
t
®
only)
®
and Practum
®
)
LEVEL
button .
l
.
User Manual Secura
®
, Quintix
Using the Balance
®
)
21
®
, Practum
®

Advertisement

Table of Contents
loading

This manual is also suitable for:

Secura3101-1sSecura26-1sSecura1103-1sSecura6101-1s

Table of Contents