Sartorius SECURA1102-1NUS User Manual page 57

Laboratory balances
Table of Contents

Advertisement

Max 220 g
4.4787
Net
+
Weight in medium
END
Max 220 g
11.86
+
To save the weight value, select
t
Remove the sample from the weighing pan and place it in the sample holder of
t
the density determination kit using forceps .
! Ensure that the sample is completely submerged in the buoyancy liquid and that
no air bubbles have formed on the sample .
y The weighing display shows the step
d = 0.0001 g
To save the weight value, select
t
y The balance calculates the density of the sample and displays this value .
d = 0.0001 g
END
To exit the function, select
t
y The application switches back to its original state . If the GLP printout function is
activated, the GLP printout is automatically printed (see "Printout" on
page 34) .
There is an online animation for this application . Click here to view it:
v
.
Weight in medium
v
.
x
.
User Manual Secura
®
, Quintix
Weighing
.
57
®