Automatic Calibration/Adjustment With Isocal - Sartorius Secura 224-1S User Manual

Laboratory balances
Table of Contents

Advertisement

Calibration and Adjustment
Max 2100 g
isoCAL
0.00
Report
Date and Time:
14-Jun-2012
Calibration function:
Internal calibration
Start calibration:
Start: isoCAL (Niv)
Calibration deviation:
Dev -0.01 g
Adjustment function:
Internal adjustment
Adjustment deviation:
Dev 0.00 g
®
76
User Manual Secura
, Quintix

8.3 Automatic Calibration/Adjustment with isoCAL

Fully automatic calibration/adjustment ensures that the balance is automatically
calibrated and adjusted when predefined time intervals or temperature values are
exceeded.
This function varies in different models:
®
Secura
/ Quintix
temperature, and each time the balance is leveled. This function can be switched
off via the system settings (see page 31).
®
In Secura
models, every adjustment process, including the identified deviations,
is documented and saved on the balance. The saved reports can be viewed via the
system settings and printed out on a PC or Sartorius lab printer (see page 31).
y If the ambient temperature has changed since the last calibration/adjustment or
the preset time interval has been exceeded, the red
d = 0.01 g
the display.
y If the
isoCAL
starts automatically.
! If the balance has not been adjusted, the prompt will be shown again 5 minutes
after the "isoCAL" status message is acknowledged.
y If the
isoCAL
manually.
t
isoCAL
To start the calibration/adjustment function manually, select the
®
(Secura
, Quintix
y Calibration starts automatically.
y A report appears on the display (Secura
The report indicates the deviations identified during isoCAL calibration.
15:18
The report also gives the result of the adjustment process.
t
Close the report window: Select
y The balance is now calibrated and adjusted.
®
®
, Practum
®
: isoCAL is automatically carried out based on time or
function is set to Automatic, the calibration/adjustment process
Info, manual start
function is set to
®
).
®
and Quintix
<
.
isoCAL
button appears on
, the function must be started
isoCAL
button
®
only).

Hide quick links: