Menu structure:

MODE

Maximum temperature (HI T)

Minimum temperature (LO T)

Maximum speed (HI R)

Minimum speed (LO R)

Fluid type (FLUI)

Menu

The type of temperature PID control arithmetic (AUTO)

Alarm and key tone (BEEP)
Calibration (CALI)

Basic guide to setting the menu options:

- Enter the menu by pressing "Menu" button (J)�
- Press the"Temp (+)" button (E) or "Temp (-)" button (F)
to change the menu options�
- Press the"Pump (+)" button (H) or "Pump (-)" button (I)
to change the value settings in the menu�
- Confirm the menu settings by pressing the "OK/Pump"
button (G)�

Operating mode (MODE):

Mode A: After power on/power failure no automatic re-
start of functions�
Mode B: After power on/power failure automatic restart of
functions, depending on previous settings�
Mode C: Set values (set in A or B) cannot be changed�
After power on/power failure automatic restart of functions,
depending on previous settings�
Mode H (H
O): If mode H is set, the fluid type (FLUI) is
2
automatically set to H
O (water)� And the temperature limit
2
is set to 5 °C to 95 °C�
The user has to adjust the safety temperature to ≤ 110 °C�
Then, the pump minimum speed (LO R) is set to 100 rpm
and the pump can be switched off separately�
Note: In mode H, the user can set the pump speed to 0
rpm via a command (Labworldsoft
function�

Menu settings

) to switch off the pump
®
A
B
C
H
AUTO 1
Proportional coefficient of PID (Kp 1)
AUTO 0
Integrate time of PID (Ti 1)
Differential time of PID (Td 1)
BEEP 0
BEEP 1
CALI 0
CALI 2
CALI 3
Maximum temperature (HI T):
The maximum adjustable value: 150 °C� This value can
be limited additionally by the selected fluids (see section
"Fluids")�
Note: The maximum value is limited by the set safe tem-
perature�
Minimum temperature (LO T):
The minimum adjustable value: -20° C
Maximum speed (HI R):
The maximum adjustable value: 3200 rpm�
Minimum speed (LO R):
The minimum adjustable value: 1000 rpm�
Default settings
activated
-
-
-
150 °C
-20 °C
3200 rpm
1000 rpm
8
activated
15�0
5�00
5�0
-
activated
activated
-
-
11