Installation
Installation at high vibration levels
NOTICE
The mounting bracket for the pipe installation of the transmitter may not be suitable for
installation environments with very high levels of vibration. In this case the user is advised
to employ more rugged methods of fixation using the threaded bottom holes directly.
If it is a remote type transmitter, it can be mounted to a pipe size DN50 (2") using the
angle bracket and retaining clip included in the delivery.
1. Screw angle bracket to bottom of transmitter.
2. Place retaining clip around pipe and slide through drill holes on angle bracket.