Print Weighing Results / Define Glp Log<P5.3 - KERN AEJ 100-4DM Operating Manual

Analysis and precision balances
Table of Contents

Advertisement

Printout example (KERN YKB-01N):
Menu item
Adjust-
ment
P5.2.1
yes
P5.2.2
yes
P5.2.3
yes
P5.2.4
yes
P5.2.5
no
P5.2.6
yes
P5.2.7
yes
P5.2.8
no
P5.2.9
no
P5.2.10
no
P5.2.11
yes
P5.2.12
no
P5.2.13
none
14.2.2
Print weighing results / define GLP log<P5.3>
Quality assurance systems require printouts of weighing results as well as of correct
adjustment of the balance stating date and time and balance identification.
To print the GLP log, press the PRINT-key.
Menu item
<P5.3.1> DATE
<P5.3.2> TIME
<P5.3.3> USER
<P5.3.4> PRODUCT
<P5.3.5> VARIABLE 1
<P5.3.6> VARIABLE 2
<P5.3.7> NET
<P5.3.8> TARE
<P5.3.9> GROSS
<P5.3.10> CURRENT RESULT
<P5.3.11> ADJUSTMENT LOG
<P5.3.12> SPECIAL PRINTOUT
AEJ-C/AES-C/PLJ-C-BA-e-1730
Protocol
Available settings / explanation
Print current date, yes / no
Print current time, yes / no
Print name of logged on user, yes / no
Print name of selected product, yes / no
Print variable 1, yes / no
For defining variable see chap. 14.4
Print variable 2, yes / no
For defining variable see chap. 14.4
Print net weight value of current weighing
Print tare value of current weighing
Print gross weight value of current weighing
Print weighing value in current unit
PRINT log of last adjustment, yes / no.
For settings of adjustment log see chap.
Options: NO / SPECIAL PRINTOUT 1 / SPECIAL PRINTOUT
2 /
SPECIAL PRINTOUT 3 / SPECIAL PRINTOUT
For creating a special printout see chap. 14.3
14.1
4 /
71

Advertisement

Table of Contents
loading