Glp/Iso Report "P5.3 Glp Print - KERN ILJ-C Operating Manual

Platform balance
Table of Contents

Advertisement

15 GLP/ISO report "P5.3 GLP PRINT"

Quality assurance systems require logs of weighing results as well as of correct
adjustment of the balance stating date and time and balance identification. The
easiest way is to have a printer connected.
The content for data output is configured specified in the "P5.3 GLP PRINT" menu.
All parameters set to "YES" will be issued.
Examples:
Definition of a standard log:
P5.3 GLP PRINT
P5.3.1
DATUM
P5.3.2
TIME
P5.3.3
USER
P5.3.4
PRODUCT
P5.3.5
VARIABLE 1
P5.3.6
VARIABLE 2
P5.3.7
TARE
P5.3.8
GROSS
P5.3.9
CURRENT
RESULT
P5.3.10
ADJUST.REP.
P5.3.11
SPECPRINT
Printout example:
---------------------------------------------------
Date
Time
User
Product
Tare
Gross
Current result
----------------------------------------------------
ILJ-C/ILS-C-BA-e-1320
YES
Issue date of measurement
YES
Issue time of measurement
YES
Issue name of logged-in user
YES
Issue name of product
NO:
Issue value of variable 1
NO:
Issue value of variable 2
YES
Tare value is issued
YES
Gross value is issued
YES
Current weighing result (net weight) is issued
NO:
The last adjustment report is issued according
to the settings for the printout of an adjustment
report.
NO:
One of the four special printouts is issued
21.11.2013
9:54:39
Kern
Flour
1000 g
2000 g
2000 g