Invoke Limits From The Product Database - KERN AEJ 100-4DM Operating Manual

Analysis and precision balances
Table of Contents

Advertisement

18.1.2 Invoke limits from the product database

Preparation
For assigning the function < SELECT PRODUCT > to a function key see chap. 17.1
Press the function key assigned to the <SELECT PRODUCT> function
(Display example F1)
Use the navigation keys to select the desired product and confirm by pressing
the PRINT-key.
The saved limits will be shown in the informational line of text (provided this
option is set for the informational line of text, see chap. 18.1).
Tare, as required, place the load on the weighing pan and start tolerance
control.
The indicators <MIN>, <OK> or< MAX> show the sample weight in relation to
the tolerance limits, see chap. 18.1.1.
• For saving a product to the product database see chap. 15.2.1.
• Return to weighing mode, see chap. 17.2.1