Channel 1 Adjustment Using DCV Calibrator

This section provides adjustment procedure using a DCV Calibrator. If a
Note
DC Power Supply is used, refer to \Channel 1 Adjustment Using DC Power
Supply".

Equipment

Setup 1/2

This setup obtains the I-V Converter scaling factors.
Figure 2-4. Channel 1 Adjustment, Setup 1/2 (using DCV Calibrator)
Perform the following steps as the program instructs.
1. Connect the 16340A furnished Triaxial Cable to the 4349B \Ch 1" connector.
2. Connect the other end of the Triaxial Cable to the Triaxial Tee Adapter (f).
2-10
Adjustments
DCV Calibrator
Multimeter
RC Box
Dual Banana-BNC(m) Cable
Triaxial Tee (m)(f)(f) Adapter
3245A
3458A
16340A
PN 11001-60001
PN 1250-1551