Setup 2/2; Channel 2, 3, And 4 Adjustments; Channel 1 Adjustment, Setup 2/2 (Using Dcv Calibrator) - Agilent Technologies 4349B Service Manual

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

3. Connect the 16340A furnished Triaxial-BNC(f) Adapter with to the Triaxial Tee Adapter (f).
4. Connect the Triaxial Tee Adapter (m) directly to the 16340A's \10 10 - 10 2 " triaxial connector.
5. Connect the 16340A's \10 10 - 10 2 " BNC connector to the calibrator dc voltage output using
the 16340A furnished BNC cable and appropriate adapter.

Setup 2/2

This setup obtains the I-V Converter input resistance data and the I-V Converter scaling factors.
Figure 2-5. Channel 1 Adjustment, Setup 2/2 (using DCV Calibrator)
Upon the program instruction, perform the following change.
1. Connect the Multimeter input and the Triaxial-BNC(f) Adapter using the Dual Banana-BNC
Cable.

Channel 2, 3, and 4 Adjustments

Perform the same setup for corresponding channel as used for Channel 1.
Channel 3 and 4 adjustments are for 4349Bs without Option 001 only.
Note
Adjustments
2-11