Measurable Flow
Breath Rate (NIV)
SpO
2
Pulse Rate
ALARM
Medium Priority
Audible Alarm
Low Priority
Audible Alarm

Electrical Information

ELECTRICAL FEATURE
AC Voltage Source
Type of protection against
Electric Shock
Degree of protection
against Electric Shock
Ingress Protection Rating
Mode of Operation
DOC-001000 Ver 6
0 - 500 l/min
Resolution 1 L/m
Tolerance ± 10%
6 - 60 BPM
Resolution 1 BPM
Tolerance ± 1 BPM
85 - 100%
Tolerance ± 3 digits
No motion or flex sensor
18 - 300 BPM
Tolerance ± 3 digits.
No motion or flex sensor
MEASURED SOUND LEVEL
45 - 50 dB(A)
45 - 50 dB(A)
SPECIFICATIONS
100 to 240 V AC, tolerance +10%/-20%,
50 to 60 Hz, max 260 W
Class II
Type BF Applied Part
Ingress Protection rating: IP22
Continuous
INDICATION
± 4 dB(A)
Measured at 1 m
± 4 dB(A)
Measured at 1 m