KERN DBS Service Manual page 11

Electronic moisture analyzer
Table of Contents

Advertisement

Setting Information Lock
Print Data
Edit Data
EEPROM
Backup
Download
Initialize All
End
DBS-SH-e-1210
Disables changing model, OEM, factory, and serial
number settings.
The balance error check function is only enabled
when settings are locked. Be sure to lock setting
information before handing over the system to the
customer.
Outputs data stored in the EEPROM to the dedicated
printer.
Edits data stored in the EEPROM using the balance
display unit.
Backs up the temperature correction coefficient,
linearity correction coefficient, and model setting status
data stored in the EEPROM to flash memory.
Writes to the EEPROM the data backed up in flash
memory.
Initializes all data saved in the EEPROM.
Ends the service mode.